PLoS ONE (Jan 2012)

Standardization of misleading immunoassay based 25-hydroxyvitamin D levels with liquid chromatography tandem-mass spectrometry in a large cohort study.

  • Ben Schöttker,
  • Eugène H J M Jansen,
  • Ulrike Haug,
  • Lutz Schomburg,
  • Josef Köhrle,
  • Hermann Brenner

DOI
https://doi.org/10.1371/journal.pone.0048774
Journal volume & issue
Vol. 7, no. 11
p. e48774

Abstract

Read online

The interest in vitamin D measurement has strongly increased in recent years. The best indicator for circulating vitamin D levels is 25-hydroxy-vitamin D (25(OH)D) which is often measured by different immunoassays. We demonstrate problems in comparability of measures by different immunoassays and the need for standardization in the context of a large population-based cohort study.25(OH)D was measured with the immunoassays Diasorin Liaison in 2006 in 5,386 women and in the context of another project with IDS-iSYS in 4,199 men in 2009-2010 (when the Diasorin Liaison was no longer available in the version utilized in 2006). Standardization was performed by re-measuring of 25(OH)D levels in 97 men and 97 women with liquid chromatography tandem-mass spectrometry (LC-MS/MS) to obtain linear regression conversion equations.Applying a 30 nmol/L cut-off value for vitamin D deficiency would have resulted in 48.3% of women and 12.1% of men with vitamin D deficiency ahead of standardization. The large gender difference was strongly attenuated after standardization of the assays with only 15.7% of women and 14.3% of men with vitamin D deficiency. Standardization on average increased the 25(OH)D levels by 10.3 nmol/L in women and decreased 25(OH)D levels by 2.9 nmol/L in men.The standardization with LC-MS/MS revealed that much of the observed gender difference was only assay-driven and the extremely high proportion of 48.3% vitamin D deficient women proved to be an exaggeration of the old version of the Diasorin-Liaison immunoassay. Standardization of 25(OH)D immunoassay results by LC-MS/MS is recommended to improve their accuracy and comparability, provided the LC-MS/MS method itself is adequately validated and standardized.