This paper describes the evaluation of a new method of sample preparation using a cold finger system with continuous water flow for rice analysis by flame atomic absorption spectrometry. The limits of detection for Cu, Fe, Mn and Zn for the proposed method were 0.36, 1.84, 2.12 and 0.16 mg kg-1, respectively. The RSDs were lower than 6.0% for all elements and the CRM analyzed showed values with 95% agreement. The proposed method is simple and safe for the proposed objective and does not require the use of mixtures of acid or special equipment for sample preparation.