International Journal of Nanomedicine (Feb 2013)

Clinical and histomorphometrical study on titanium dioxide-coated external fixation pins

  • Koseki H,
  • Asahara T,
  • Shida T,
  • Yoda I,
  • Horiuchi H,
  • Baba K,
  • Osaki M

Journal volume & issue
Vol. 2013, no. default
pp. 593 – 599

Abstract

Read online

Hironobu Koseki,1 Tomohiko Asahara,1 Takayuki Shida,1 Itaru Yoda,1 Hidehiko Horiuchi,1 Koumei Baba,2 Makoto Osaki11Department of Orthopedic Surgery, Graduate School of Medicine, Nagasaki University, 2Industrial Technology Center of Nagasaki, Nagasaki, JapanBackground: Pin site infection is the most common and significant complication of external fixation. In this work, the efficacy of pins coated with titanium dioxide (TiO2) for inhibition of infection was compared with that of stainless steel control pins in an in vivo study.Methods: Pins contaminated with an identifiable Staphylococcus aureus strain were inserted into femoral bone in a rat model and exposed to ultraviolet A light for 30 minutes. On day 14, the animals were sacrificed and the bone and soft tissue around the pin were retrieved. The clinical findings and histological findings were evaluated in 60 samples.Results: Clinical signs of infection were present in 76.7% of untreated pins, but in only 36.7% of TiO2-coated pins. The histological bone infection score and planimetric rate of occupation for bacterial colonies and neutrophils in the TiO2-coated pin group were lower than those in the control group. The bone-implant contact ratio of the TiO2-coated pin group was significantly higher (71.4%) than in the control pin group (58.2%). The TiO2 was successful in decreasing infection both clinically and histomorphometrically.Conclusion: The photocatalytic bactericidal effect of TiO2 is thought to be useful for inhibiting pin site infection after external fixation.Keywords: titanium dioxide, external fixation, bactericidal activity, Staphylococcus aureus