Journal of Medical Physics (Jan 2022)
In vivo Dosimetry for dose verification of total skin electron beam therapy using gafchromic® EBT3 film dosimetry
Abstract
Background and Purpose: Total skin electron beam therapy (TSEBT) is an important skin-directed radiotherapeutic procedure done in the treatment of cutaneous T-cell lymphomas, namely, mycosis fungoides (MF). This procedure is usually done at larger source-to-surface distances with the patient standing on a rotatory platform. As the patient has to stand in different positions without any rigid immobilization devices, there are chances that the total skin may not get uniformly irradiated which could lead to nonuniform dose distributions. Therefore, all the necessary arrangements should be made to evaluate the dose for different regions of the skin using suitable in vivo dosimeters at the radiotherapy centers offering these treatments. This study aimed to evaluate the consistency between the delivered and planned doses in vivo during TSEBT using Gafchromic EBT3 film dosimetry. Materials and Methods: The surface dose for the six MF patients treated for TSEBT at our hospital from 2018 to 2022 was measured and evaluated. 2 cm × 2 cm Gafchromic® EBT3 films were used to measure skin dose at reference body positions of clinical interest. All the patients were treated with the modified Stanford technique. The irradiated film strips were analyzed for the dose using the IMRT OmniPro software. The doses at respective positions were expressed as mean dose ± standard deviation and the deviation was calculated as the percentage of the prescribed dose. Results: One hundred and fifty-four Gafchromic® EBT3 film strips irradiated on six TSEBT patients showed a maximum dose variation of 2.00 ± 0.14 Gy, in the central body regions. The dose variation in the peripheral areas such as hands and ears was larger. A variation of 2 ± 0.32 Gy was observed on the hands and ears. The uniformity of the dose delivered to maximum body parts was within −7% and +16% for the peripheral areas like hands. The American Association of Physicists in Medicine recommends a dose uniformity of 8% and 4% in the vertical and horizontal patient plane for direct incident beam; however, for oblique incidences like in the modified Stanford technique, the dose variation is about 15%. Conclusion: In vivo dosimetry using Gafchromic EBT3 film dosimetry for TSEBT yields objective data to find the under or overdose regions. That can be useful to provide quality treatment, especially when treatments tend to be as complex as TSEBT.
Keywords