Indian Journal of Ophthalmology (Jan 2006)

Probable edge defect in Acrysof single-piece intraocular lens

  • Shah Urmi

Journal volume & issue
Vol. 54, no. 4
pp. 277 – 278

Abstract

Read online

The purpose of this article is to report an edge defect in single-piece hydrophobic acrylic intraocular lens (IOL) observed during a routine phacoemulsification procedure. The chip was successfully removed intraoperatively with a McPherson′s forceps. However, six months postoperatively patient complained of a disturbing glare due to posterior capsular folds originating from the site of IOL defect. Patient was advised, however refused to undergo Nd:Yag laser capsulotomy to relieve the symptoms. This case highlights the necessity to check each foldable IOL under the microscope before loading it into the cartridge, for possible edge defects.

Keywords