Effect of kilovoltage and quality reference mAs on CT-based attenuation correction in 177Lu SPECT/CT imaging: a phantom study

  • Maikol Salas-Ramirez,
  • Julian Leube,
  • Michael Lassmann,
  • Johannes Tran-Gia

DOI
https://doi.org/10.1186/s40658-024-00622-6
Journal volume & issue
Vol. 11, no. 1
pp. 1 – 16

Abstract

Read online

Abstract Introduction CT-based attenuation correction (CT-AC) plays a major role in accurate activity quantification by SPECT/CT imaging. However, the effect of kilovoltage peak (kVp) and quality-reference mAs (QRM) on the attenuation coefficient image (μ-map) and volume CT dose index (CTDIvol) have not yet been systematically evaluated. Therefore, the aim of this study was to fill this gap and investigate the influence of kVp and QRM on CT-AC in 177Lu SPECT/CT imaging. Methods Seventy low-dose CT acquisitions of an Electron Density Phantom (seventeen inserts of nine tissue-equivalent materials) were acquired using various kVp and QRM combinations on a Siemens Symbia Intevo Bold SPECT/CT system. Using manufacturer reconstruction software, 177Lu μ-maps were generated for each CT image, and three low-dose CT related aspects were examined. First, the μ-map-based attenuation values (μ measured) were compared with theoretical values (μ theoretical). Second, changes in 177Lu activity expected due to changes in the μ-map were calculated using a modified Chang method. Third, the noise in the μ-map was assessed by measuring the coefficient of variation in a volume of interest in the homogeneous section of the Electron Density Phantom. Lastly, two phantoms were designed to simulate attenuation in four tissue-equivalent materials for two different source geometries (1-mL and 10-mL syringes). 177Lu SPECT/CT imaging was performed using three different reconstruction algorithms (xSPECT Quant, Flash3D, STIR), and the SPECT-based activities were compared against the nominal activities in the sources. Results The largest relative errors between μ measured and μ theoretical were observed in the lung inhale insert (range: 18%-36%), while it remained below 6% for all other inserts. The resulting changes in 177Lu activity quantification were -3.5% in the lung inhale insert and less than -2.3% in all other inserts. Coefficient of variation and CTDIvol ranged from 0.3% and 3.6 mGy (130 kVp, 35 mAs) to 0.4% and 0.9 mGy (80 kVp, 20 mAs), respectively. The SPECT-based activity quantification using xSPECT Quant reconstructions outperformed all other reconstruction algorithms. Conclusion This study shows that kVp and QRM values in low-dose CT imaging have a minimum effect on quantitative 177Lu SPECT/CT imaging, while the selection of low values of kVp and QRM reduce the CTDIvol.

Keywords