SICOT-J (Jan 2023)
Analysis of an explanted dual mobility cup after 21 years, parabolic wear pattern: A case report
Abstract
Case: A dual mobility cup was implanted in 1983 in a 43-year woman. After 31 years of normal function, we analysed the explanted materials with modern techniques. Conclusion: Volumetric wears of the small and large articulations of the dual-mobility construct were similar. For the first time, we demonstrated that the dual-mobility liner underwent parabolic (not linear) wear during the period of implantation.
Keywords