Journal of Anaesthesiology Clinical Pharmacology (Jan 2012)

Complete bacterial filter blockage by a plastic film

  • Jiapeng Huang,
  • Kishin Dodwani,
  • Paul Simon,
  • Jing Zhou

DOI
https://doi.org/10.4103/0970-9185.94902
Journal volume & issue
Vol. 28, no. 2
pp. 230 – 231

Abstract

Read online

We describe a case where the expiratory bacterial filter was completely blocked by a plastic membrane. Significantly increased airway pressures and inability to deliver adequate tidal volume to the patient were experienced. The cause of machine failure could not be identified until a complete preanesthesia machine manual check was performed. All machines and circuits should be checked manually, even with the availability of new and advanced self-tests in machines. Backup ventilation equipments and familiarity with signs of expiratory and inspiratory limb obstruction should not be underestimated.