Sensors (Mar 2015)

An Improved Electrochemical Aptasensor for Chloramphenicol Detection Based on Aptamer Incorporated Gelatine

  • Ezat Hamidi-Asl,
  • Freddy Dardenne,
  • Ronny Blust,
  • Karolien De Wael

DOI
https://doi.org/10.3390/s150407605
Journal volume & issue
Vol. 15, no. 4
pp. 7605 – 7618

Abstract

Read online

Because of the biocompatible properties of gelatine and the good affinity of aptamers for their targets, the combination of aptamer and gelatine type B is reported as promising for the development of biosensing devices. Here, an aptamer for chloramphenicol (CAP) is mixed with different types of gelatine and dropped on the surface of disposable gold screen printed electrodes. The signal of the CAP reduction is investigated using differential pulse voltammetry. The diagnostic performance of the sensor is described and a detection limit of 1.83 × 10−10 M is found. The selectivity and the stability of the aptasensor are studied and compared to those of other CAP sensors described in literature.

Keywords