口腔疾病防治 (Jan 2021)

The significance and parameter standards of the implant primary stability

  • XU Jing

DOI
https://doi.org/10.12016/j.issn.2096⁃1456.2021.01.001
Journal volume & issue
Vol. 29, no. 1
pp. 1 – 10

Abstract

Read online

The nature, significance, parameters, influencing factors and testing of implant primary stability were studied by a literature review. Primary stability is a kind of anchorage force at the interface between the implant and bone, and it is merely mechanical. The significance of primary stability is to keep the implant unmovable so that the new bone can grow undisturbed on the surface of the implant without interference from fibrous tissue. The implant is finally bound to the bone by osseointegration. The most common assessments of primary stability are insertion torque (IT), the implant stability quotient (ISQ) of the resonance frequency analysis (RFA) and Periotest. IT is more commonly used to directly imply initial stability. At present, no consensus has been reached regarding the concrete parameters of primary stability to predict osseointegration. Implant osseointegration could be developed through all phases of primary stability. However, the excessive primary stability would cause mini-bone fractures, followed by bone necrosis at the interface and the final failure of implantation. Primary stability is influenced by three factors: implant design, bone condition of alveolar bone, and surgical technique. Under the condition of a lack of primary stability and immediate implantation, there may be the possibility of successful osseointegration. Therefore, it is necessary to re-examine the accuracy of the current elaboration on the primary stability. It is related directly to whether the clinic can choose the appropriate implant treatment path.

Keywords