BMJ Paediatrics Open (Nov 2024)
What CPAP to use in the delivery room? Bench comparison of two methods to provide continuous positive airways pressure in neonates
Abstract
Background Continuous positive airway pressure (CPAP) is a recommended first-line therapy for infants with respiratory distress at birth. Resuscitation devices incorporating CPAP delivery can have significantly different imposed resistances affecting airway pressure stability and work of breathing.Aim To compare CPAP performance of two resuscitation devices (Neopuff T-piece resuscitator and rPAP) in a neonatal lung model simulating spontaneous breathing effort at birth.Methods The parameters assessed were variation in delivered pressures (∆P), tidal volume (VT), inspiratory effort (model pressure respiratory muscle (PRM)) and work of breathing (WOB). Two data sequences were required with Neopuff and one with rPAP: (1) set PRM with changes in VT and (2) constant VT (preterm 6 mL, term 22 mL) with increased effort. Data were collected at CPAP settings of 5, 7 and 9 cmH2O using a 1 kg preterm (Compliance: 0.5 mL/cmH2O) and 3.5 kg term (1.0 mL/cmH2O) model.Results 2298 breaths were analysed (760 rPAP, 795 Neopuff constant VT, 743 Neopuff constant PRM). With CPAP at 9 cmH2O and set VT the mean ∆P (cmH2O) rPAP vs Neopuff 1.1 vs 5.6 (preterm) and 1.9 vs 13.4 (term), WOB (mJ) 4.6 vs 6.1 (preterm) and 35.3 vs 44.5 (term), and with set PRM mean VT (ml) decreased to 6.2 vs 5.2 (preterm) and 22.3 vs 17.5 (term) p<0.001. Similar results were found at pressures of 5 and 7 cmH2O.Conclusion rPAP had smaller pressure swings than Neopuff at all CPAP levels and was thus more pressure stable. WOB was higher with Neopuff when VT was held constant. VT reduced with Neopuff when respiratory effort was constant.