Respiratory Research (Mar 2024)
Respiratory entrainment related reverse triggering in mechanically ventilated children
Abstract
Abstract Background The underlying pathophysiological pathways how reverse triggering is being caused are not fully understood. Respiratory entrainment may be one of these mechanisms, but both terms are used interchangeably. We sought to characterize reverse triggering and the relationship with respiratory entrainment among mechanically ventilated children with and without acute lung injury. Methods We performed a secondary phyiology analysis of two previously published data sets of invasively mechanically ventilated children < 18 years with and without lung injury mechanically ventilated in a continuous or intermittent mandatory ventilation mode. Ventilator waveforms, electrical activity of the diaphragm measured with surface electromyography and oesophageal tracings were analyzed for entrained and non-entrained reverse triggered breaths. Results In total 102 measurements (3110 min) from 67 patients (median age 4.9 [1.8 ; 19,1] months) were analyzed. Entrained RT was identified in 12 (12%) and non-entrained RT in 39 (38%) recordings. Breathing variability for entrained RT breaths was lower compared to non-entrained RT breaths. We did not observe breath stacking during entrained RT. Double triggering often occurred during non-entrained RT and led to an increased tidal volume. Patients with respiratory entrainment related RT had a shorter duration of MV and length of PICU stay. Conclusions Reverse triggering is not one entity but a clinical spectrum with different mechanisms and consequences. Trial registration Not applicable.
Keywords