Revista Brasileira de Cineantropometria e Desempenho Humano (Feb 2023)

Validation of the Trackhealth Physical Activity Monitor

  • Ricardo Ansaloni de Oliveira,
  • Jeffer Eidi Sasaki,
  • Natália Lujan Ferraz,
  • Álvaro Ribeiro Gomes de Oliveira,
  • Pedro Henrique Ataides de Moraes,
  • Jair Sindra Virtuoso Júnior

DOI
https://doi.org/10.1590/1980-0037.2022v24e83854
Journal volume & issue
Vol. 24, no. 1

Abstract

Read online

This is a quantitative methodological study for the validation of a research instrument. It aimed to validate the data from the TrackHealth accelerometry device. The sample consisted of 30 adult individuals of both sexes selected by convenience who met the inclusion and exclusion criteria. The physical activity monitors used for the research protocol were the ActiGraph® wGT3X-BT triaxial accelerometer and the TrackHealth accelerometer (TH). The activity protocol consisted of 4 (four) activities (walking at 4.8 and 6.4 km h¹ and running at 9.7 and 12 km h¹) performed in the laboratory, on an Ibramed treadmill, lasting 5 (five) minutes at each stage. A difference was found between the raw acceleration data of the two devices, however the TrackHealth device showed higher sensitivity at speeds of 4.8 and 6.4 km/h, and a high level of agreement (2.7-2.8%) at the initial speeds of the magnitude vectors. However, there is still a need for improvement in the functioning of the device, so that TrackHealth can be commercialized.