BACKGROUND: Reading capsule endoscopy (CE) is time-consuming. The Express View (EV) (IntroMedic, Seoul, Korea) software was designed to shorten CE video reading. Our primary aim was to evaluate the diagnostic accuracy of EV in detecting significant small-bowel (SB) lesions. We also compared the reading times with EV mode and standard reading (SR). METHODS: 126 patients with suspected SB bleeding and/or suspected neoplasia were prospectively enrolled and underwent SB CE (MiroCam®1200, IntroMedic, Seoul, Korea). CE evaluation was performed in standard and EV mode. In case of discrepancies between SR and EV readings, a consensus was reached after reviewing the video segments and the findings were re-classified. RESULTS: The completion rate of SB CE in our cohort was 86.5% and no retention occurred. The per-patient analysis of sensitivity, specificity, positive predictive value, negative predictive value, and diagnostic accuracy of EV compared to SR were 86%, 86%, 90%, 81%, and 86%, respectively, before consensus. After consensus, they increased to 97%, 100%, 100%, 96%, and 98%, respectively. The median reading time with SR and EV was 71 min (range 26–340) and 13 min (range 3–85), respectively (p < 0.001). CONCLUSIONS: The new-generation EV shows high diagnostic accuracy and significantly reduces CE reading times.