Comparison of Bilevel Volume Guarantee and Pressure-Regulated Volume Control Modes in Preterm Infants
Abstract
The present study aimed to compare the bilevel volume guarantee (VG) and pressure-regulated volume control (PRVC) modes of the GE® Carescape R860 model ventilator and test the safety and feasibility of these two modes in preterm neonates. Infants who were less than 30 weeks of gestational age were included. After randomization, initial ventilator settings were adjusted for each patient. After the first 2 h of ventilation, the patients were switched to the other ventilator mode for 2 h. The ventilator parameters, vital signs, and blood gas values were evaluated. The study included a total of 28 patients, 14 in the PRVC group and 14 in the bilevel VG group. The mean birth weight was 876 g (range: 530–1170) and the mean gestational age was 26.4 weeks (range: 24–29). The patients’ peak inspiratory pressure (PIP2 and PIP3) was lower after ventilation in bilevel VG mode than in PRVC mode (13 vs. 14 cmH2O, respectively; paired samples t-test, p = 0.008). After 2 h of bilevel VG ventilation, the mean heart rate decreased from 149/min to 140/min (p = 0.001) and the oxygen saturation increased from 91% to 94% (p = 0.01). Both the PRVC and bilevel VG modes of GE ventilators can be used safely in preterm infants, and bilevel VG mode was associated with more favorable early clinical findings. Studies including more patients and comparing with other modes will clarify and provide further evidence on this subject.