Sensors (Dec 2005)

Amperometric Detection of Nitric Oxide with Microsensor in the Medium of Seawater and Its Applications

  • Cai Lin,
  • Zhen-Zhen Wu,
  • Chun-Ying Liu,
  • Zheng-Bin Zhang,
  • Lei Xing

DOI
https://doi.org/10.3390/s5120537
Journal volume & issue
Vol. 5, no. 12
pp. 537 – 545

Abstract

Read online

Amperometric detection of nitric oxide with ISO-NOPMC NO microsensors(WPI) is systemically studied in the six media including seawater. The linear range of themicrosensor for NO was from 10-6 to 10-9 mol/l and the detection limit was4.2×10-10 mol/L(S/N=2). With this method, we provide preliminary evidence that NOproduction could be a general attribute of marine alga (Haeterosigma akashiwo).Experiment conducted with inhibitor of uncoupler 2,4-DNP (2,4-dinitropheno) revealed thatNR (nitrate reductase) activity is responsible for NO formation.

Keywords