EPJ Techniques and Instrumentation (Apr 2017)

The aerosol impact spectrometer: a versatile platform for studying the velocity dependence of nanoparticle-surface impact phenomena

  • Brian D. Adamson,
  • Morgan E. C. Miller,
  • Robert E. Continetti

DOI
https://doi.org/10.1140/epjti/s40485-017-0037-6
Journal volume & issue
Vol. 4, no. 1
pp. 1 – 17

Abstract

Read online

Abstract A new apparatus designed to accelerate/decelerate and study the surface impact phenomena of charged aerosols and nanoparticles over a wide range of mass-to-charge (m/z) ratios and final velocities is described. A nanoparticle ion source coupled with a linear electrostatic trap configured as an image charge detection (ICD) mass spectrometer allows determination of the mass-to-charge ratio and the absolute charge and mass of single nanoparticles. A nine-stage linear accelerator/decelerator is used to fix the final velocity of the nanoparticles, and in the results reported here the coefficient of restitution for polystyrene latex spheres (PSLs) impacting on silicon is measured using ICD techniques. To enable this apparatus to study a wide range of m/z, the data acquisition system uses a transient digitizer interfaced to a field-programmable gate array module that allows real time calculation of m/z and determination of the pulse sequence for the linear accelerator/decelerator. Electrospray ionization of a colloidal suspension of PSL spheres of 510 and 990 nm has been used to demonstrate acceleration and deceleration of charged nanoparticles and the resolution of the apparatus. Measurements of the coefficient of restitution for PSLs on silicon over the range 10-400 m/s are consistent with previous studies.

Keywords