Sensors (Sep 2021)

A Low Excitation Working Frequency Capacitively Coupled Contactless Conductivity Detection (C<sup>4</sup>D) Sensor for Microfluidic Devices

  • Yuchen He,
  • Qiang Huang,
  • Yu He,
  • Haifeng Ji,
  • Tao Zhang,
  • Baoliang Wang,
  • Zhiyao Huang

DOI
https://doi.org/10.3390/s21196381
Journal volume & issue
Vol. 21, no. 19
p. 6381

Abstract

Read online

In this work, a new capacitively coupled contactless conductivity detection (C4D) sensor for microfluidic devices is developed. By introducing an LC circuit, the working frequency of the new C4D sensor can be lowered by the adjustments of the inductor and the capacitance of the LC circuit. The limits of detection (LODs) of the new C4D sensor for conductivity/ion concentration measurement can be improved. Conductivity measurement experiments with KCl solutions were carried out in microfluidic devices (500 µm × 50 µm). The experimental results indicate that the developed C4D sensor can realize the conductivity measurement with low working frequency (less than 50 kHz). The LOD of the C4D sensor for conductivity measurement is estimated to be 2.2 µS/cm. Furthermore, to show the effectiveness of the new C4D sensor for the concentration measurement of other ions (solutions), SO42− and Li+ ion concentration measurement experiments were also carried out at a working frequency of 29.70 kHz. The experimental results show that at low concentrations, the input-output characteristics of the C4D sensor for SO42− and Li+ ion concentration measurement show good linearity with the LODs estimated to be 8.2 µM and 19.0 µM, respectively.

Keywords