IHJ Cardiovascular Case Reports (Jan 2018)

Add-on pacing/sensing lead for successful management of “Defibrillator Shock Storm”: Repair of a Make and Break Circuit

  • Naveen Garg,
  • Deep Chandh Raja

Journal volume & issue
Vol. 2, no. 1
pp. 61 – 63

Abstract

Read online

A case of “ICD storm” precipitated by lead noise generation due to fracture in the sensing/pacing circuit component of the ICD lead was successfully repaired by excluding the damaged sensing/pacing circuit component of the ICD lead and putting an additional parallel ventricular lead connected with the existing generator.