Journal of Otology (Oct 2021)
Objective frequency-specific hearing thresholds definition for medicolegal purposes in case of occupational NIHL: ASSR outperforms CERA
Abstract
Audiological use of the 40 Hz-ASSR (auditory steady state responses) could be valuable for objectively estimating the frequency-specific threshold in adults undergoing an expertise examination for medicolegal and/or compensation purposes. The present prospective study was set up to clarify the relationship between the thresholds obtained by cortical evoked response audiometry (CERA) and by 40 Hz-ASSR, in the same ears, within a large homogeneous sample of 164 subjects (328 ears) with NIHL and well documented exposure to noise. All these subjects claimed financial compensation for occupational NIHL, and there was a suspicion of exaggeration of the reported NIHLs. ASSR thresholds show a good correlation with the CERA thresholds. However, a systematic shift is noticed, ASSR thresholds being on average (1–2 – 3 kHz) 4.38 dB lower (i.e. showing less hearing loss) than CERA thresholds. Moreover, the binaural multiple ASSR technique allows a considerable time gain when compared to the CERA.