BMC Anesthesiology (Jul 2023)

Development and assessment of the performance of a shared ventilatory system that uses clinically available components to individualize tidal volumes

  • David M. Hannon,
  • Tim Jones,
  • Jack Conolly,
  • Conor Judge,
  • Talha Iqbal,
  • Atif Shahzad,
  • Michael Madden,
  • Frank Kirrane,
  • Peter Conneely,
  • Brian H. Harte,
  • Martin O’Halloran,
  • John G. Laffey

DOI
https://doi.org/10.1186/s12871-023-02200-2
Journal volume & issue
Vol. 23, no. 1
pp. 1 – 11

Abstract

Read online

Abstract Objectives To develop and assess a system for shared ventilation using clinically available components to individualize tidal volumes. Design Evaluation and in vitro validation study Setting Ventilator shortage during the SARS-CoV-2 pandemic. Participants The team consisted of physicians, bioengineers, computer programmers, and medical technology professionals. Methods Using clinically available components, a system of ventilation consisting of two ventilatory limbs was assembled and connected to a ventilator. Monitors for each limb were developed using open-source software. Firstly, the effect of altering ventilator settings on tidal volumes delivered to each limb was determined. Secondly, the impact of altering the compliance and resistance of one limb on the tidal volumes delivered to both limbs was analysed. Experiments were repeated three times to determine system variability. Results The system permitted accurate and reproducible titration of tidal volumes to each limb over a range of ventilator settings and simulated lung conditions. Alteration of ventilator inspiratory pressures, of respiratory rates, and I:E ratio resulted in very similar tidal volumes delivered to each limb. Alteration of compliance and resistance in one limb resulted in reproducible alterations in tidal volume to that test lung, with little change to tidal volumes in the other lung. All tidal volumes delivered were reproducible. Conclusions We demonstrate the reliability of a shared ventilation system assembled using commonly available clinical components that allows titration of individual tidal volumes. This system may be useful as a strategy of last resort for Covid-19, or other mass casualty situations, where the need for ventilators exceeds supply.