Journal of Ophthalmology (Jan 2019)
Application of 14-MHz Ultrasonography with Tissue Harmonic Imaging to Determine Posterior Capsule Integrity in Traumatic Cataract
Abstract
Purpose. To report the application of 14-MHz ultrasonography with tissue harmonic imaging (14 MHz + THI) to determine the integrity of the posterior capsule (PC) in traumatic cataract (TC). Methods. Patients with TC who were scheduled to undergo cataract extraction and whose PC could not be observed by slit lamp examination were included in the study. The status of the PC was determined by 14 MHz + THI before cataract extraction and confirmed during surgery. The results regarding PC integrity obtained from 14 MHz + THI and intraoperative direct observation were compared. Result. The study enrolled 52 eyes of 52 patients (49 men and 3 women), with a mean age of 42.15 years ± 11.23 (SD). The nature of the trauma was blunt (3 eyes) or sharp (49 eyes). The 14 MHz + THI method showed 21 PCs to be intact and 31 to have ruptured before cataract surgery. During surgery, 23 PCs were observed to be intact, while 29 PCs were ruptured. 27 PCs were ruptured and 19 were intact, as determined by the two methods. The 14 MHz + THI observations were consistent with the intraoperative observations of the PC (kappa = 0.764), with no significant difference between the two methods (P=0.687). The sensitivity, specificity, and accuracy of 14 MHz + THI for observation of the PC were 93.10%, 82.60%, and 88.46%, respectively. Conclusion. The 14 MHz + THI method can accurately reveal the integrity of the PC in TC. It has important clinical value in the selection of cataract surgery methods and the prediction of complications during TC surgery.