PLoS ONE (Jan 2011)

Implant size and fixation mode strongly influence tissue reactions in the CNS.

  • Jonas Thelin,
  • Henrik Jörntell,
  • Elia Psouni,
  • Martin Garwicz,
  • Jens Schouenborg,
  • Nils Danielsen,
  • Cecilia Eriksson Linsmeier

DOI
https://doi.org/10.1371/journal.pone.0016267
Journal volume & issue
Vol. 6, no. 1
p. e16267

Abstract

Read online

The function of chronic brain machine interfaces depends on stable electrical contact between neurons and electrodes. A key step in the development of interfaces is therefore to identify implant configurations that minimize adverse long-term tissue reactions. To this end, we here characterized the separate and combined effects of implant size and fixation mode at 6 and 12 weeks post implantation in rat (n = 24) cerebral cortex. Neurons and activated microglia and astrocytes were visualized using NeuN, ED1 and GFAP immunofluorescence microscopy, respectively. The contributions of individual experimental variables to the tissue response were quantified. Implants tethered to the skull caused larger tissue reactions than un-tethered implants. Small diameter (50 µm) implants elicited smaller tissue reactions and resulted in the survival of larger numbers of neurons than did large diameter (200 µm) implants. In addition, tethering resulted in an oval-shaped cavity, with a cross-section area larger than that of the implant itself, and in marked changes in morphology and organization of neurons in the region closest to the tissue interface. Most importantly, for implants that were both large diameter and tethered, glia activation was still ongoing 12 weeks after implantation, as indicated by an increase in GFAP staining between week 6 and 12, while this pattern was not observed for un-tethered, small diameter implants. Our findings therefore clearly indicate that the combined small diameter, un-tethered implants cause the smallest tissue reactions.