The Journal of Engineering (Jun 2017)

Time-division polynomial pre-distorter for linearisation of 1.5 T MRI power amplifier

  • Ming Hui,
  • Meng Zhang,
  • Xingang Zhang,
  • Xiaowei Zhu,
  • Xiupu Zhang,
  • Jiali Cheng

DOI
https://doi.org/10.1049/joe.2017.0195

Abstract

Read online

A time-division polynomial (TDP) model is proposed for modelling and linearising a 1.5 T magnetic resonance imaging (MRI) power amplifier (PA) with strong non-linearity in high input signal dynamic range. In order to demonstrate the merit of this non-linear model, a 64 dBm 1.5 T MRI PA (63.89 MHz) and two different Sinc-pulse signals are used in modelling and linearisation measurements. The TDP is compared with the conventional non-memory polynomial (NMP) and no digital pre-distortion for the 1.5 T MRI PA, which is driven by test signal with 2 ms time length and 2% duty cycle. The proposed TDP leads to up to 9 dB improvement in the normalised mean square error compared with the NMP in two different test signals. More importantly, TDP illustrates significantly better reduction in amplitude modulation/amplitude modulation (AM/AM) and amplitude modulation/phase modulation (AM/PM) conversion compared with the NMP.

Keywords