MethodsX (Jun 2024)

An improved sample container cup and its effect on NIRS of tobacco smoke for quality stability estimation

  • Juan Huo,
  • Huaiqi Li,
  • Changtong Lu

Journal volume & issue
Vol. 12
p. 102478

Abstract

Read online

This paper describes a design of an improved self-made Bruker NIR cup and analyzes the effect of the equipment modification to fit the Cambridge filter pad, which enhances experimental efficiency and reduces operational complexity. A self-made NIR cup based on the classical NIR cup is designed to speed up the operation process and reduce the experiment's time cost. To estimate the effect of this equipment modification, the NIR spectra from the classical sample cup and the new self-made cup are compared and analyzed. Furthermore, the quality evaluation results from NIR data of the two cups are also compared according to a distance metric chemometrics method, which shows quality analytical values between these two cups are approaching each other while the experiment efficiency is improved. • This paper introduces a newdesign of a self-made container cup improved from the Bruker's traditional sample container cup to better fit the filter pad and improve the experiment efficiency and convenience. • This paper also analyzes the effect of this container cup change by comparing the NIR spectra before and after modification.

Keywords