Sensors (Feb 2012)

Development of an Integrated Chip for Automatic Tracking and Positioning Manipulation for Single Cell Lysis

  • Chao-Wang Young,
  • Chyung Ay,
  • Jia-Ling Hsieh

DOI
https://doi.org/10.3390/s120302400
Journal volume & issue
Vol. 12, no. 3
pp. 2400 – 2413

Abstract

Read online

This study adopted a microelectromechanical fabrication process to design a chip integrated with electroosmotic flow and dielectrophoresis force for single cell lysis. Human histiocytic lymphoma U937 cells were driven rapidly by electroosmotic flow and precisely moved to a specific area for cell lysis. By varying the frequency of AC power, 15 V AC at 1 MHz of frequency configuration achieved 100% cell lysing at the specific area. The integrated chip could successfully manipulate single cells to a specific position and lysis. The overall successful rate of cell tracking, positioning, and cell lysis is 80%. The average speed of cell driving was 17.74 μm/s. This technique will be developed for DNA extraction in biomolecular detection. It can simplify pre-treatment procedures for biotechnological analysis of samples.