Physical Properties of Modern Reciprocal Endodontic Systems and Fatigue Failure Testing in Simulated Clinical Conditions
Abstract
Instrument fracture ranks among the most crucial complications during the endodontic treatment of a tooth. In order to better understand the practical limits of the instrument, the relation between the cyclic fatigue resistance and physical properties such as hardness, modulus of elasticity, creep and surface roughness were explored. Cyclic fatigue testing in an artificial root canal at intracanal temperature, nanoindentation and 3D microscopy were used for evaluation of four commonly used thermomechanically treated NiTi endodontic instruments (Unicone Plus 6/025, Unicone 6/025, Reciproc Blue R25 and WaveOne Gold Primary). Cyclic fatigue results were analyzed using the Kruskal–Wallis, Mann–Whitney and Bonferroni corrections. The wear resistance of Unicone 6/025 instruments was significantly lower compared to all other instruments (p p > 0.05). These results are in correlation with measurements of local mechanical properties (hardness, elastic modulus and their ratios H/E and H3/E2). Even though surface roughness, area of cross-section and shape of instruments are important factors affecting instruments behavior, thermal processing appears to be the most important.