International Journal of Endocrinology (Jan 2009)

Clinical Evaluation of Subcutaneous Lactate Measurement in Patients after Major Cardiac Surgery

  • Martin Ellmerer,
  • Martin Haluzik,
  • Jan Blaha,
  • Jaromir Kremen,
  • Stepan Svacina,
  • Andreas Plasnik,
  • Dimas Ikeoka,
  • Manfred Bodenlenz,
  • Lukas Schaupp,
  • Johannes Plank,
  • Thomas R. Pieber

DOI
https://doi.org/10.1155/2009/390975
Journal volume & issue
Vol. 2009

Abstract

Read online

Minimally invasive techniques to access subcutaneous adipose tissue for glucose monitoring are successfully applied in type1 diabetic and critically ill patients. During critical illness, the addition of a lactate sensor might enhance prognosis and early intervention. Our objective was to evaluate SAT as a site for lactate measurement in critically ill patients. In 40 patients after major cardiac surgery, arterial blood and SAT microdialysis samples were taken in hourly intervals. Lactate concentrations from SAT were prospectively calibrated to arterial blood. Analysis was based on comparison of absolute lactate concentrations (arterial blood vs. SAT) and on a 6-hour lactate trend analysis, to test whether changes of arterial lactate can be described by SAT lactate. Correlation between lactate readings from arterial blood vs. SAT was highly significant (r2=0.71, P<.001). Nevertheless, 42% of SAT lactate readings and 35% of the SAT lactate trends were not comparable to arterial blood. When a 6-hour stabilization period after catheter insertion was introduced, 5.5% of SAT readings and 41.6% of the SAT lactate trends remained incomparable to arterial blood. In conclusion, replacement of arterial blood lactate measurements by readings from SAT is associated with a substantial shortcoming in clinical predictability in patients after major cardiac surgery.