PLoS ONE (Jan 2021)

Towards distortion-free imaging of the eye.

  • Phillip Bedggood,
  • Andrew Metha

DOI
https://doi.org/10.1371/journal.pone.0252876
Journal volume & issue
Vol. 16, no. 6
p. e0252876

Abstract

Read online

The high power of the eye and optical components used to image it result in "static" distortion, remaining constant across acquired retinal images. In addition, raster-based systems sample points or lines of the image over time, suffering from "dynamic" distortion due to the constant motion of the eye. We recently described an algorithm which corrects for the latter problem but is entirely blind to the former. Here, we describe a new procedure termed "DIOS" (Dewarp Image by Oblique Shift) to remove static distortion of arbitrary type. Much like the dynamic correction method, it relies on locating the same tissue in multiple frames acquired as the eye moves through different gaze positions. Here, the resultant maps of pixel displacement are used to form a sparse system of simultaneous linear equations whose solution gives the common warp seen by all frames. We show that the method successfully handles torsional movement of the eye. We also show that the output of the previously described dynamic correction procedure may be used as input for this new procedure, recovering an image of the tissue that is, in principle, a faithful replica free of any type of distortion. The method could be extended beyond ocular imaging, to any kind of imaging system in which the image can move or be made to move across the detector.