Frontiers in Physiology (Oct 2018)

A Calibration Device to Compare Body Plethysmographs Among Pediatric Lung Function Laboratories

  • Bruno Demoulin,
  • Iulia Ioan,
  • Iulia Ioan,
  • Claude Duvivier,
  • Claude Bonabel,
  • Cyril Schweitzer,
  • Cyril Schweitzer,
  • Cyril Schweitzer,
  • François Marchal,
  • François Marchal,
  • Silvia Demoulin-Alexikova,
  • Silvia Demoulin-Alexikova

DOI
https://doi.org/10.3389/fphys.2018.01408
Journal volume & issue
Vol. 9

Abstract

Read online

Multi-center studies in specific airway resistance have shown significant inter laboratory variability. Comparison of plethysmographic equipment using a lung model easily transportable from one site to another should be of help to international normative studies. A resistor made of parallel capillary tubes – insuring adequate linearity within 1 L/sec – was connected to a glass bottle. Thermal time constants were measured while the bottle was empty and while stuffed with steel wool. In the latter, isothermal condition was estimated to occur only at very low frequency (around 0.01 Hz) and gas compression was polytropic up to 0.6 Hz. With the empty analog, adiabatic gas compression was estimated to occur at frequencies ≥0.2 Hz, and more accurate volume estimation was obtained. The empty analog volume and specific resistance measured in a body plethysmograph on different days indicated within 5% accuracy as well as intersession repeatability. It is concluded that a physical analog built out of simple material provides accurate measurements of specific resistance. The apparatus should be of help to compare plethysmographic equipments from different laboratories.