Journal of Clinical Medicine (Jun 2023)

Prospects of Endovenous Laser Ablation (EVLA) Standardization—Mid-Term Results of a Four-Zone Dosimetry Guiding Tool for 1940 nm Laser

  • Abhay Setia,
  • Slobodan Dikic,
  • Sahit Demhasaj,
  • Thomas Schmitz-Rixen,
  • Ronald Sroka,
  • Claus-Georg Schmedt

DOI
https://doi.org/10.3390/jcm12134313
Journal volume & issue
Vol. 12, no. 13
p. 4313

Abstract

Read online

Background: Heterogeneity regarding dosimetry and reporting of endovenous laser ablation (EVLA) mandates the development of a standardized protocol. This study presents the mid-term results of EVLA with 1940 nm-laser and radial-fibre, supported by a four-zone dosimetry tool. Materials and methods: Four anatomical dosimetry zones for great saphenous veins (GSV) and two for small saphenous veins (SSV) were defined with set power levels. Zone-1G (4 W) extended from the inguinal ligament to the apex of femoral triangle, Zone-2G (4 W) from the apex of femoral triangle to the upper border of patella. Zone-3G (3 W) from the patella to the tibial tuberosity. Zone-4G (2 W) extended from the tibial tuberosity to the ankle. Zone-1S from the sapheno-popliteal junction to the tibial tuberosity. Zone-2S from the tibial tuberosity to the ankle. Power was increased by 1 W for veins >10 mm and decreased by 1 W when fibre sticking was encountered. Pullback-velocity was max. 1 mm/s. Results: A total of 152 consecutive patients (185 procedures) were recruited. Mean follow-up time was 11.9 months. Mean linear endovenous energy density for GSV was Zone-1G:42 J/cm, Zone-2G:33 J/cm, Zone-3G:27 J/cm, Zone-4G:22 J/cm, Zone-1S:34 J/cm, Zone-2S:27 J/cm. Occlusion rates were 98.9% (1-month) and 93.7% (12-months). Complications at 1 M were low, namely laser-induced paraesthesia (LIP) 2.2% and endovenous heat-induced thrombosis (EHIT) 1.6%. Persistent LIP (12 M) was observed in 0.5%. Conclusion: The proposed four-zone guiding tool is a step towards standardizing dosimetry and documentation for EVLA with 1940 nm. This strategy shows good mid-term results with minimal complications. Long-term follow-up and application in further centres are necessary to prove its reproducibility. Such a guiding tool could improve the ability to analyse, compare and review different EVLA wavelengths and fibre types.

Keywords