Bioengineering (Jun 2024)
Assessing the Impact of PRESERFLO MicroShunt on Intraocular Pressure in Porcine Eyes Ex Vivo Using Infusion Pump System
Abstract
To evaluate the effectiveness of the PRESERFLO MicroShunt (PFM) in reducing intraocular pressure (IOP) ex vivo in porcine eyes using an infusion pump system and to simulate various IOP conditions, In this study, porcine eyes received increasing flows between 2 and 20 μL/min. IOP measurements were taken under conditions with and without the PFM [PFM (+) and PFM (−), respectively]. In the PFM (−) group, IOP increased from 7.4 mmHg to 46.3 mmHg as the flow rate increased from 2 μL/min to 20 μL/min. The rate of IOP reduction (%ΔIOP) rose with increasing flow rates, although the absolute IOP values achieved with the PFM insertion also increased. The correlation between IOPs in the PFM (−) conditions and the %ΔIOP was modeled as %ΔIOP = 22.4 Ln [PFM(−) IOP] − 41.7. According to this equation, IOP reduction by PFM insertion is 0% at IOPs of 6.4 mmHg or lower. IOP reductions of 10%, 20%, 30%, and 40% were observed when the pre-insertion IOPs were 10.1, 15.7, 24.6, and 38.4 mmHg, respectively. Achievable post-insertion IOP levels of ≤21 mmHg, ≤18 mmHg, ≤15 mmHg, and ≤12 mmHg corresponded to the initial IOPs of 33 mmHg, 26 mmHg, 20 mmHg, and 14.8 mmHg, respectively. In conclusion, the PFM effectively reduced IOP within a specific range of IOP values in an ex vivo experimental system. In clinical situations, the PFM is unlikely to be effective at low IOP levels. At higher levels, the PFM reduces IOP, but it may be insufficient to achieve the target IOP.
Keywords