Journal of Imaging (Feb 2022)

Mixed-Reality-Assisted Puncture of the Common Femoral Artery in a Phantom Model

  • Christian Uhl,
  • Johannes Hatzl,
  • Katrin Meisenbacher,
  • Lea Zimmer,
  • Niklas Hartmann,
  • Dittmar Böckler

DOI
https://doi.org/10.3390/jimaging8020047
Journal volume & issue
Vol. 8, no. 2
p. 47

Abstract

Read online

Percutaneous femoral arterial access is daily practice in a variety of medical specialties and enables physicians worldwide to perform endovascular interventions. The reported incidence of percutaneous femoral arterial access complications is 3–18% and often results from suboptimal puncture location due to insufficient visualization of the target vessel. The purpose of this proof-of-concept study was to evaluate the feasibility and the positional error of a mixed-reality (MR)-assisted puncture of the common femoral artery in a phantom model using a commercially available navigation system. In total, 15 MR-assisted punctures were performed. Cone-beam computed tomography angiography (CTA) was used following each puncture to allow quantification of positional error of needle placements in the axial and sagittal planes. Technical success was achieved in 14/15 cases (93.3%) with a median axial positional error of 1.0 mm (IQR 1.3) and a median sagittal positional error of 1.1 mm (IQR 1.6). The median duration of the registration process and needle insertion was 2 min (IQR 1.0). MR-assisted puncture of the common femoral artery is feasible with acceptable positional errors in a phantom model. Future studies should aim to measure and reduce the positional error resulting from MR registration.

Keywords