Physics and Imaging in Radiation Oncology (Apr 2018)

Dosimetric end-to-end tests in a national audit of 3D conformal radiotherapy

  • Joerg Lehmann,
  • Andrew Alves,
  • Leon Dunn,
  • Maddison Shaw,
  • John Kenny,
  • Stephanie Keehan,
  • Jeremy Supple,
  • Francis Gibbons,
  • Sophie Manktelow,
  • Chris Oliver,
  • Tomas Kron,
  • Ivan Williams,
  • Jessica Lye

Journal volume & issue
Vol. 6
pp. 5 – 11

Abstract

Read online

Background and purpose: Independent dosimetry audits improve quality and safety of radiation therapy. This work reports on design and findings of a comprehensive 3D conformal radiotherapy (3D-CRT) Level III audit. Materials and methods: The audit was conducted as onsite audit using an anthropomorphic thorax phantom in an end-to-end test by the Australian Clinical Dosimetry Service (ACDS). Absolute dose point measurements were performed with Farmer-type ionization chambers. The audited treatment plans included open and half blocked fields, wedges and lung inhomogeneities. Audit results were determined as Pass Optimal Level (deviations within 3.3%), Pass Action Level (greater than 3.3% but within 5%) and Out of Tolerance (beyond 5%), as well as Reported Not Scored (RNS). The audit has been performed between July 2012 and January 2018 on 94 occasions, covering approximately 90% of all Australian facilities. Results: The audit pass rate was 87% (53% optimal). Fifty recommendations were given, mainly related to planning system commissioning. Dose overestimation behind low density inhomogeneities by the analytical anisotropic algorithm (AAA) was identified across facilities and found to extend to beam setups which resemble a typical breast cancer treatment beam placement. RNS measurements inside lung showed a variation in the opposite direction: AAA under-dosed a target beyond lung and over-dosed the lung upstream and downstream of the target. Results also highlighted shortcomings of some superposition and convolution algorithms in modelling large angle wedges. Conclusions: This audit showed that 3D-CRT dosimetry audits remain relevant and can identify fundamental global and local problems that also affect advanced treatments. Keywords: Radiation therapy audit, End-to-end test, Level III, Australian Clinical Dosimetry Service, 3D-CRT