Scientific Reports (Jun 2024)
The appropriate sample-handling procedure for measuring the plasma β-amyloid level using a fully automated immunoassay
Abstract
Abstract Plasma β-amyloid (Aβ) assays are a promising tool for Alzheimer’s disease diagnosis in clinical practice. To obtain reliable results, establishing an appropriate sample-handling procedure for each analytical platform is warranted. This study proposes an appropriate sample-handling procedure using HISCL analyzer by elucidating the individual/combined effects of pre-analytical parameters on plasma Aβ42/Aβ40 levels. We investigated the effects of various pre-analytical parameters, including storage times for whole blood, plasma, and freezing conditions, on plasma Aβ42/Aβ40 levels, and confirmed if these values met the acceptable criteria. Plasma Aβ42/Aβ40 levels were acceptable in all conditions. We determined our protocol by confirming that plasma Aβ42/Aβ40 levels remained acceptable when combining pre-analytical parameters. We established an appropriate sample-handling protocol that ensures reliable measurement of plasma Aβ42/Aβ40 levels using HISCL analyzer. We believe the Aβ assay, with our protocol, shows promise for aiding AD diagnosis in clinical settings.