Scientific Reports (Jun 2024)

The appropriate sample-handling procedure for measuring the plasma β-amyloid level using a fully automated immunoassay

  • Kengo Ishiki,
  • Kazuto Yamashita,
  • Shunsuke Watanabe,
  • Masahiro Miura,
  • Junko Kawahira,
  • Yuji Arimatsu,
  • Kana Kawasaki,
  • Shigeki Iwanaga,
  • Toshiyuki Sato

DOI
https://doi.org/10.1038/s41598-024-65264-1
Journal volume & issue
Vol. 14, no. 1
pp. 1 – 10

Abstract

Read online

Abstract Plasma β-amyloid (Aβ) assays are a promising tool for Alzheimer’s disease diagnosis in clinical practice. To obtain reliable results, establishing an appropriate sample-handling procedure for each analytical platform is warranted. This study proposes an appropriate sample-handling procedure using HISCL analyzer by elucidating the individual/combined effects of pre-analytical parameters on plasma Aβ42/Aβ40 levels. We investigated the effects of various pre-analytical parameters, including storage times for whole blood, plasma, and freezing conditions, on plasma Aβ42/Aβ40 levels, and confirmed if these values met the acceptable criteria. Plasma Aβ42/Aβ40 levels were acceptable in all conditions. We determined our protocol by confirming that plasma Aβ42/Aβ40 levels remained acceptable when combining pre-analytical parameters. We established an appropriate sample-handling protocol that ensures reliable measurement of plasma Aβ42/Aβ40 levels using HISCL analyzer. We believe the Aβ assay, with our protocol, shows promise for aiding AD diagnosis in clinical settings.