Measurement of Photo-Neutron Dose from an 18-MV Medical Linac Using a Foil Activation Method in View of Radiation Protection of Patients
Abstract
High-energy linear accelerators are increasingly used in the medical field. However, the unwanted photo-neutrons can also be contributed to the dose delivered to the patients during their treatments. In this study, neutron fluxes were measured in a solid water phantom placed at the isocenter 1-m distance from the head of an 18-MV linac using the foil activation method. The produced activities were measured with a calibrated well-type Ge detector. From the measured fluxes, the total neutron fluence was found to be (1.17 ± 0.06) × 107 n/cm2 per Gy at the phantom surface in a 20 × 20 cm2 X-ray field size. The maximum photo-neutron dose was measured to be 0.67 ± 0.04 mSv/Gy at dmax = 5 cm depth in the phantom at isocenter. The present results are compared with those obtained for different field sizes of 10 × 10cm2, 15 × 15cm2, and 20 × 20cm2 from 10-, 15-, and 18-MV linacs. Additionally, ambient neutron dose equivalents were determined at different locations in the room and they were found to be negligibly low. The results indicate that the photo-neutron dose at the patient position is not a negligible fraction of the therapeutic photon dose. Thus, there is a need for reduction of the contaminated neutron dose by taking some additional measures, for instance, neutron absorbing-protective materials might be used as aprons during the treatment.