Diagnostics (Nov 2021)

Validation of a Monte Carlo Modelling Based Dosimetry of Extraoral Photobiomodulation

  • Anna N. Yaroslavsky,
  • Amy F. Juliano,
  • Ather Adnan,
  • Wayne J. Selting,
  • Tyler W. Iorizzo,
  • James D. Carroll,
  • Stephen T. Sonis,
  • Christine N. Duncan,
  • Wendy B. London,
  • Nathaniel S. Treister

DOI
https://doi.org/10.3390/diagnostics11122207
Journal volume & issue
Vol. 11, no. 12
p. 2207

Abstract

Read online

An in vivo validation study was performed to confirm the accuracy of extraoral photobiomodulation therapy (PBMT) dosimetry determined by modelling. The Monte Carlo technique was utilized to calculate the fluence rate and absorbed power of light delivered through multi-layered tissue. Optical properties used during Monte Carlo simulations were taken from the literature. Morphological data of four study volunteers were acquired using magnetic resonance imaging (MRI) scans. Light emitting diode (LED) coupled to a power meter were utilized to measure transmitted power through each volunteer’s cheek, in vivo. The transmitted power determined by Monte Carlo modelling was compared to the in vivo measurements to determine the accuracy of the simulations. Experimental and simulation results were in good agreement for all four subjects. The difference between the mean values of the measured transmission was within 12% from the respective transmission obtained using Monte Carlo simulations. The results of the study indicate that Monte Carlo modelling is a robust and reliable method for light dosimetry.