Scientific Reports (Dec 2023)

Application of double low-dose mode in left atrial-pulmonary venous computed tomography angiography

  • Changjiang Zhang,
  • Wei Zhang,
  • Kaihu Shi,
  • Jingya Chen

DOI
https://doi.org/10.1038/s41598-023-48973-x
Journal volume & issue
Vol. 13, no. 1
pp. 1 – 7

Abstract

Read online

Abstract This study adopted a 256-slice iCT scanner with the double low-dose mode in left atrial-pulmonary venous computed tomography angiography (CTA) and explored its effect on image quality. 120 patients were included and randomly classified into the Observation group and Control group. Patients in the Control group underwent routine left atrial CTA, while patients in the Observation group performed a double low-dose mode. Other scanning parameters were consistent in the two groups. The Full model-based iterative reconstruction (MBIR) technique was applied to fulfill image reconstruction in observation group. Continuous variables, ordered categorical variables were analyzed by statistical test. The CT values of left atrial in the Observation group were significantly higher than those in the Control group. The exposure doses (ED) and iodine intake were lower in the Observation group, as compared to the Control group. The left atrial-pulmonary venous CTA with the 256-slice iCT scanner in a double low-dose mode can reduce the ED of radiation and iodine contrast while providing high quality images. Comparatively, the ED in the Observation group was reduced by 13% compared with the control, and the iodine intake was reduced by approximately 33%.