BMC Anesthesiology (Aug 2018)

A comparison of two hyperangulated video laryngoscope blades to direct laryngoscopy in a simulated infant airway: a bicentric, comparative, randomized manikin study

  • Marc Kriege,
  • Nina Pirlich,
  • Thomas Ott,
  • Eva Wittenmeier,
  • Frank Dette

DOI
https://doi.org/10.1186/s12871-018-0580-y
Journal volume & issue
Vol. 18, no. 1
pp. 1 – 9

Abstract

Read online

Abstract Background In infants, securing the airway is time-critical because of anatomical and physiological differences related to airway management in children less than 1 year old. The aim of this study was to compare the time to ventilation using two different hyperangulated video laryngoscope blades with the time to ventilation via conventional direct laryngoscopy in a normal airway [NA] and in a simulated difficult airway [DA]. Methods This study was a comparative, bicentric, open-label, randomized controlled evaluation. An infant high-fidelity simulator (SimBaby™; Laerdal® Medical, Stavanger, Norway) was used, and two scenarios were proposed, as follows: NA and DA evoked with tongue edema and cervical collar. After theoretical and practical briefing, each participant compared in the two airway scenarios the novel King Vision™ Pediatric aBlade (KV) (Ambu® A/S, Bad Nauheim, Germany) video laryngoscope and the C-MAC™ D-blade Ped (DB) (Karl Storz® SE & Co. KG, Tuttlingen, Germany) video laryngoscope to conventional laryngoscopy using the Miller Blade (MiB) and the Macintosh Blade (MaB) in a random sequence. Results Eighty physicians (65 AN and 15 PCCM staff) were included. In the NA scenario, the median [IQR] time to successful time to ventilation (TTV) was significantly shorter for the KV at 13 s [12–15 s] than for the MaB at 14.5 s [13–16 s], DB at 14.5 s [13–16] and MiB at 16 s [14–19] (p < 0.001). In DA, the KV also shortened TTV to 14 s [13–16], whereas TTV was 23 s with the MaB [20–26], 19 s with the DB [16–21], and 27 s with the MiB [22–31] (p < 0.001). There were no differences in first-pass intubation success rates (FPAs) between hyperangulated blades and direct laryngoscopes in NA. In DA, the hyperangulated blades enabled 92 (DB) to 100% (KV) FPAs compared with 65 (MiB) to 76% (MaB) for conventional laryngoscopy (p < 0.001). Conclusion Video laryngoscopes with hyperangulated blades were associated with shorter TTV in normal and difficult infant airway situations. The higher FPAs of hyperangulated blades in DA may avoid desaturations and decrease adverse events in pediatric airway management.

Keywords