Scientific Reports (Sep 2021)

Comparison of SpO2 and heart rate values on Apple Watch and conventional commercial oximeters devices in patients with lung disease

  • Leonardo Zumerkorn Pipek,
  • Rafaela Farias Vidigal Nascimento,
  • Milena Marques Pagliarelli Acencio,
  • Lisete Ribeiro Teixeira

DOI
https://doi.org/10.1038/s41598-021-98453-3
Journal volume & issue
Vol. 11, no. 1
pp. 1 – 7

Abstract

Read online

Abstract Lung diseases have high mortality and morbidity, with an important impact on quality of life. Hypoxemic patients are advised to use oxygen therapy to prolong their survival, but high oxygen saturation (SpO2) levels can also have negative effects. Pulse oximeters are the most common way to assess oxygen levels and guide medical treatment. This study aims to assess whether wearable devices can provide precise SpO2 measurements when compared to commercial pulse oximeters. This is a cross-section study with 100 patients with chronic obstructive pulmonary disease and interstitial lung disease from an outpatient pneumology clinic. SpO2 and heart rate data were collected with an Apple Watch Series 6 (Apple) and compared to two commercial pulse oximeters. The Bland–Altman method and interclass correlation coefficient were used to compare their values. We observed strong positive correlations between the Apple Watch device and commercial oximeters when evaluating heart rate measurements (r = 0.995, p 0.05). Apple Watch 6 is a reliable way to obtain heart rate and SpO2 in patients with lung diseases in a controlled environment.