The Journal of Engineering (Dec 2019)
In-situ calibration technology of liquid level with precise laser ranging
Abstract
Here, in order to calibrate the magnetostrictive liquid-level meter in situ, a set of in-situ calibration device for magnetostrictive liquid-level meter is designed. The structure and working principle of the device are described, and the experimental test and error analysis are carried out for the parts of the in-situ calibration device. Finally, according to the modern instrument and error theory, the accuracy of the calibration device is assessed, and the field calibration experiment of magnetostrictive liquid level gauge is carried out. The theoretical analysis and field experiment results show that the calibrated device can meet the in-situ calibration requirements of the magnetostrictive liquid-level meter with a range of 0–6 m and a precision of 1 mm.
Keywords