Nuclear Engineering and Technology (Apr 2014)

SPECTRUM WEIGHTED RESPONSES OF SEVERAL DETECTORS IN MIXED FIELDS OF FAST AND THERMAL NEUTRONS

  • SANG IN KIM,
  • INSU CHANG,
  • BONG HWAN KIM,
  • JANG LYUL KIM,
  • JUNG IL LEE

DOI
https://doi.org/10.5516/NET.08.2013.029
Journal volume & issue
Vol. 46, no. 2
pp. 273 – 280

Abstract

Read online

The spectrum weighted responses of various detectors were calculated to provide guidance on the proper selection and use of survey instruments on the basis of their energy response characteristics on the neutron fields. To yield the spectrum weighted response, the detector response functions of 17 neutron-measuring devices were numerically folded with each of the produced calibration neutron spectra through the in-house developed software ‘K-SWR’. The detectors’ response functions were taken from the IAEA Technical Reports Series No. 403 (TRS-403). The reference neutron fields of 21 kinds with 2 spectra groups with different proportions of thermal and fast neutrons have been produced using neutrons from the 241Am-Be sources held in a graphite pile, a bare 241Am-Be source, and a DT neutron generator. Fluence-average energy (Eave) varied from 3.8 MeV to 16.9 MeV, and the ambient-dose-equivalent rate [H*(10)/h] varied from 0.99 to 16.5 mSv/h.

Keywords