A Novel Instrumentation Circuit for Electrochemical Measurements
Abstract
In this paper, a novel signal processing circuit which can be used for the measurement of H<sup>+</sup> ion and urea concentration is presented. A potentiometric method is used to detect the concentrations of H<sup>+</sup> ions and urea by using H<sup>+</sup> ion-selective electrodes and urea electrodes, respectively. The experimental data shows that this measuring structure has a linear pH response for the concentration range within pH 2 and 12, and the dynamic range for urea concentration measurement is in the range of 0.25 to 64 mg/dL. The designed instrumentation circuit possesses a calibration function and it can be applied to different sensing electrodes for electrochemical analysis. It possesses the advantageous properties of being multi-purpose, easy calibration and low cost.
Keywords