Sensors (Nov 2006)

Design of a Flow-through Polarographic Sensor Based on Metal Films for Determining N-nitrosodiethanolamine Levels in Rabbit Biological Fluids

  • Yuan-Zhi Lan,
  • Hung-Chang Hsia,
  • Lai-Hao Wang

DOI
https://doi.org/10.3390/s6111555
Journal volume & issue
Vol. 6, no. 11
pp. 1555 – 1567

Abstract

Read online

The construction and characterization of a flow-through polarographic detectorfor catalyzing the electroreduction of N-nitrosodiethanolamine (NDELA), is discussed. Theflow-through cell is equipped with a gold wire electrode (a thin mercury film deposited on agold substance). The response is evaluated with respect to substance diameter, length,concentration of modifying film, operating potential, supporting electrolyte and pH, andflow rate in the DC mode. The system allows the determination of N-nitrosodiethanolaminein rabbit biological fluids with relatively inexpensive equipment.