Journal of the Korean Society of Radiology (Sep 2023)

Variability of Transrectal Shear Wave Elastography in a Phantom Model

  • Jihyun Lee,
  • Seong Kuk Yoon,
  • Jin Han Cho,
  • Hee Jin Kwon,
  • Dong Won Kim,
  • Jun Woo Lee

DOI
https://doi.org/10.3348/jksr.2023.0051
Journal volume & issue
Vol. 84, no. 5
pp. 1110 – 1122

Abstract

Read online

Purpose This study aimed to assess the variability of transrectal shear wave elastography (SWE) using a designed phantom. Materials and Methods In a phantom, the SWE values were examined by two radiologists using agarose and emulsion silicone of different sizes (1, 2, and 3 cm) and shapes (round, cubic) at three depths (1, 2, and 3 cm), two region of interest (ROI) and locations (central, peripheral) using two ultrasound machines (A, B from different vendors). Variability was evaluated using the coefficient of variation (CV). Results The CVs decreased with increasing phantom size. Significant changes in SWE values included; agarose phantom at 3 cm depth (p 0.9). Conclusion The phantom size, depth, and different machines affected the variability of transrectal SWE.

Keywords