Current Directions in Biomedical Engineering (Oct 2021)

Particle Image Velocimetry for in vitro characterization of Paravalvular Leakage of TAVR-sealing concepts

  • Höing Samuel,
  • Borowski Finja,
  • Oldenburg Jan,
  • Illner Sabine,
  • Öner Alper,
  • Stiehm Michael,
  • Schmitz Klaus-Peter,
  • Kaule Sebastian

DOI
https://doi.org/10.1515/cdbme-2021-2170
Journal volume & issue
Vol. 7, no. 2
pp. 668 – 671

Abstract

Read online

Paravalvular leakage (PVL), defined as the leakage between the aortic annulus and a transcatheter aortic valve replacement (TAVR), is verifiably associated with short- and long-term clinical outcome, especially with increased mortality. Therefore, with the ambition to reduce or even prevent PVL of next generation TAVR, it is necessary to extend the hemodynamic understanding of PVL.

Keywords