Indian Journal of Dental Sciences (Jan 2019)

Assessment of efficacy and maintenance of light-curing units in dental offices across Punjab: A clinical survey

  • Rajinder Bansal,
  • Manu Bansal,
  • Shilpa Walia,
  • Loveena Bansal,
  • Karanvir Singh,
  • Ridhi Aggarwal

DOI
https://doi.org/10.4103/IJDS.IJDS_63_18
Journal volume & issue
Vol. 11, no. 1
pp. 42 – 45

Abstract

Read online

Aims and Objectives: Curing units in dental offices across Punjab are analyzed for a practical purpose. Materials and Methods: One thousand light-curing units regularly used by the dentists in dental offices across the state of Punjab were examined for their output intensity. Various factors include as follows: type, power of light-curing unit, material molecule formation on tip (YES/NO), diameter of guide tip, and frequency of bulb replacement. A radiometer, a magnifying glass, and a Vernier caliper were used. Data were collected and analyzed. Results: About 75.80% of dentists use light-emitting diode (LED) lights, while 24.20% utilize halogen lamps to cure dental composites. 36.60% of light-curing units of the state had light intensity below 300 mW/cm2, out of which 61.60% were quartz–tungsten–halogen (QTH) and 28.60% LED light-curing units. 17.40% of light-curing units recorded light intensity between 301 and 400 mW/cm2. 46% of light-curing units registered an output intensity of >400 mW/cm2. 79.60% of light units showed material particle buildup on light-curing tips. 62.40% of dentists never replaced the bulbs of their light-curing units. Two-third of dental the practitioners avoided infection control barriers on the tips of curing units. Conclusions: There is a deficit of knowledge among dental practitioners in Punjab regarding care of light-curing units. Dental light-curing units should be regularly checked and infection control remedies should be opted.

Keywords