International Journal of Ophthalmology (Dec 2019)

Impact of corneal parameters on intraocular pressure measurements in different tonometry methods

  • Aleksandra Zakrzewska,
  • Marta P. Wiącek,
  • Anna Machalińska

DOI
https://doi.org/10.18240/ijo.2019.12.06
Journal volume & issue
Vol. 12, no. 12
pp. 1853 – 1858

Abstract

Read online

AIM: To evaluate the impact of central corneal thickness (CCT) and corneal curvature on intraocular pressure (IOP) measurements performed by three different tonometers. METHODS: IOP in 132 healthy eyes of 66 participants was measured using three different tonometry techniques: Goldmann applanation tonometer (GAT), Pascal dynamic contour tonometer (DCT), and ICare rebound tonometer (RT). CCT and corneal curvature were assessed. RESULTS: In healthy eyes, DCT presents significantly higher values of IOP than GAT (17.34±3.69 and 15.27±4.06 mm Hg, P<0.0001). RT measurements are significantly lower than GAT (13.56±4.33 mm Hg, P<0.0001). Compared with GAT, DCT presented on average 2.51 mm Hg higher values in eyes with CCT<600 μm and 0.99 mm Hg higher results in eyes with CCT≥600 μm. The RT results were lower on average by 1.61 and 1.95 mm Hg than those obtained by GAT, respectively. Positive correlations between CCT in eyes with CCT<600 μm were detected for all IOP measurement techniques, whereas a similar relationship was not observed in eyes with thicker corneas. A correlation between IOP values and keratometry in the group with CCT<600 μm was not detected with any of the tonometry methods. In thicker corneas, a positive correlation was found for GAT and mean keratometry values (R=0.369, P=0.005). CONCLUSION: The same method should always be chosen for routine IOP control, and measurements obtained by different methods cannot be compared. All analysed tonometry methods are dependent on CCT; thus, CCT should be taken into consideration for both diagnostics and monitoring.

Keywords