Antimicrobial Resistance and Infection Control (Jan 2019)

Comparison of GeneXpert MRSA/SA ETA assay with semi-quantitative and quantitative cultures and nuc gene-based qPCR for detection of Staphylococcus aureus in endotracheal aspirate samples

  • Jasmine Coppens,
  • Liesbet Van Heirstraeten,
  • Alexey Ruzin,
  • Li Yu,
  • Leen Timbermont,
  • Christine Lammens,
  • Veerle Matheeussen,
  • Michael McCarthy,
  • Philippe Jorens,
  • Margareta Ieven,
  • Samir Kumar-Singh,
  • Herman Goossens,
  • Surbhi Malhotra-Kumar

DOI
https://doi.org/10.1186/s13756-018-0460-8
Journal volume & issue
Vol. 8, no. 1
pp. 1 – 7

Abstract

Read online

Abstract Introduction Staphylococcus aureus (S. aureus) is a common cause of ventilator-associated pneumonia. Rapid and accurate detection of lower respiratory tract colonization and/or infection with S. aureus may inform targeted preventive and therapeutic strategies. To investigate this, we compared semi-quantitative (SQ)-culture results from 79 endotracheal aspirates (ETA) collected from mechanically-ventilated patients, to two culture and two non-culture-based methods for detection of S. aureus. Methods ETA analyzed by routine SQ-culture on blood and colistin-nalidixic-acid agar was compared to: (i) quantitative (Q-) culture on chromogenic COLOREX™ Staph aureus; (ii) enrichment in brain-heart-infusion broth followed by plating on blood agar and COLOREX™; (iii) nuc-based TaqMan qPCR, and (iv) GeneXpert MRSA/SA ETA assay. Results Of the 79 ETA samples analyzed by SQ-culture, 39 samples were positive, and 40 negative for S. aureus. Two samples negative for S. aureus by SQ-culture were, however, S. aureus-positive by the other four methods and were considered positive. Appending these two samples as positive in the SQ-culture results, sensitivities−specificities for Q-culture, enrichment-culture, TaqMan qPCR and GeneXpert were 100–95, 100–92, 100–53% and 100% − 100, respectively. The lower specificities of Q-culture, enrichment-culture, and TaqMan qPCR was because of their higher sensitivities, although TaqMan qPCR also detected S. aureus-specific extracellular DNA. Conclusion This first evaluation of the GeneXpert MRSA/SA ETA assay with ETA samples found it to be highly sensitive, specific, user-friendly (hands-on time ~ 5 min.), and rapid (~ 66 min. assay time). Where this equipment is not available, we recommend implementing more sensitive culture-based methods for improved S. aureus detection in ETA samples.

Keywords