Applied Sciences (Mar 2022)

Biomechanical Analysis of Patient-Specific Temporomandibular Joint Implant and Comparison with Natural Intact Jaw Bone Using Finite Element Method

  • Anubhav Tiwari,
  • Vijay Kumar Gupta,
  • Rakesh Kumar Haldkar,
  • Ivan A. Parinov

DOI
https://doi.org/10.3390/app12063003
Journal volume & issue
Vol. 12, no. 6
p. 3003

Abstract

Read online

The purpose of this study is to design a patient-specific TMJ implant and study its behaviour under different loading conditions compared with natural intact TMJ. There are several diseases, which affect the proper growth and function of TMJ, and in some cases, TMJ injury results from accidents. To repair the TMJ, temporomandibular joint replacement or TJR surgery is performed. In this work, CT-scan data of the skull and mandible region with broken condylar head were used to study the biomechanical behaviour of the intact mandible and customized TMJ prostheses in order to design a patient-specific total TMJ implant. The customized TMJ implant was virtually studied under simulated loading conditions using finite element method (FEM) in ANSYS Workbench and then compared to the intact jaw-mandible for the combinations of two different biocompatible material models. It is observed that the natural TMJ has a higher deformation value as compared to the patient-specific TMJ implant due to the lower mechanical strength of bone relative to the Ti-6Al-4V and Co-Cr alloy. Hence, we can conclude that the designed custom TMJ implant is safe for the patient from the point of design perspective.

Keywords